DIATON tonometer

IOP MEASURING DEVICE

Diaton tonometer is the only one device for intraocular pressure measuring through the upper eyelid. It completely meets modern diagnostic equipment requirements.

pictures: image_1   pictures: image_2

Fields of diaton tonometer application  

  • ophthalmology (including children's)
  • general medical practice
  • optometry
  • neurology
  • Method for measuring the intraocular pressure through the eyelid and device for realizing the same are protected with the Patent of Russia № 2123798, United States Patent № US 6,394,954 В1 and Patent of Japan №3593314.  

    Currently, transpalpebral scleral tonometry has no alternative and is the most optimum method when carrying out prophylactic medical examination of the population and in complicated clinical cases when the use of classical tonometry methods is impossible.

    Innovative approach to IOP measurement opens up wide clinical possibilities and unquestionable advantages to the doctor and to the patient:

    pictures: image_3

    for the doctor:

    • accurate tonometry result,
    • wide clinical possibilities,
    • image of modern doctor and health center,
    • quickness and efficiency (no disposables, IOP measurement takes several seconds),
    • spectrum service spreading.

    for the patient:  

    • safety,
    • quickness,
    • comfort,
    • painless,
    • no necessity to take off contact lenses
     

    Transpalpebral "diaton" tonometer is effective and irreplaceable in various situations: 
    • screening examinations of the patients
    • IOP control during selection of adequate medicines
    • IOP measuring in the presence in a patient of chronical conjunctivitis, erosions, edema and cornea dimness
    • IOP measuring in patients after corneal surgeries
    • ophthalmotone day monitoring
    • IOP measuring in immobilized patients and in children
    • IOP measuring during contact correction (lenses are not taken out).

     

    FeaturesdiatonGoldmann
    Tonometer
    Shiotz
    Tonometer
    Air-jetТonopeni-care
    No contact with the cornea +          
    Portability +   +   + +
    Displays independence from cornea's crookedness.
    +
             
    Digital IOP indication +     + + +
    Measurement in sitting position + +   + + +
    Measurement in reclining position +   +   +  
    Short-time measurement +     +   +
    Sterilization is not required +     +    
    Anesthesia is not required +     +    

    Comparison tests of diaton tonometer and Goldman tonometer are the evidence of high reliability of diaton tonometer displays.

    Specifications

    Measurement range, mm Hg 5-60
    Measurement error Limit of the admissible measurement error in the range, not more:
    from 5 to 20 mm Hg - ±2 mm Hg;
    from 20 to 60 mm Hg - ±10%
    The time of a single measurement, s, not more 3
    Supply voltage, V 3
    Number of measurements using one battery set, not less 1500
    Service life, not less 5
    Weight, g 89
    Dimensions, mm, not more 174 х 26 х 20

     