Comparative characteristics of Diaton-tonometry results
while changing the tonometer position on the eyelid

Alla R. Illarionova
Academic Group of A.P.Nesterov, academician of Russian Academy of Medical Sciences (Moscow)

#Name, age IOP GoldmanDiaton-tonometry
Correct Position (Variant 1) Measuring through the Limb (Variant 2) Measuring through the Cornea (Variant 3)
1PEM, 4730
16
27
15
21
13
18
10
2VLY, 5229
10
27
14
14
6
11
6
3KNK151266
4SIK, 58 8
14
12
16
8
15
6
10
5IAK, 46 13
14
15
14
11
10
8,0
8,0
6BPP18
17
16
16
12
14
9,0
7,0
7ChPN, 6714
13
15
15
13
13
8,0
6,0
8DYG, 4812
14
11,0
12,0
9,0
9,0
7,0
9,0
9FPO, 6140
20
24
22
23
18
19
13
10UAP, 7711
17
14
17
10
13
8,0
10
11FEI, 58 36
13
27
14
22
10
18
9,0
12OOO40
18
27
19
25
16
19
8,0
13TYB, 67 16
15
14
14
10
10
6,0
7,0

Analysis of Measurement Results

 Var1Var2 Var3Var4
MEAN 18,5217,1613,249,84
MEDIAN 1515138
SD 9,060537145,249444425,133874434,2296572
VALID_N 25252525
MIN case 1-6181166
MAX case 1-6140272519
R0,870,610,52




While analyzing the results of IOP measuring using one and the same device by one and the same investigator, the tonometer position on the eyelid being changed, the mean values closeness was registered only when the position of tip on the eyeball was correct.

The sufficient correlation of the results was observed only in this case.

While moving the tip lower in the limb area or on the cornea the result is significantly lower and is not true.

Ryazan State Instrument-making Enterprise &tm;