Comparative characteristics of Diaton-tonometry results while changing the tonometer position on the eyelid
Alla R. Illarionova
Academic Group of A.P.Nesterov, academician of Russian Academy of Medical Sciences (Moscow)
| # | Name, age |
IOP Goldman | Diaton-tonometry |
| Correct Position (Variant 1) |
Measuring through the Limb (Variant 2) |
Measuring through the Cornea (Variant 3) |
| 1 | PEM, 47 | 30
16 | 27 15 | 21 13 | 18 10 |
| 2 | VLY, 52 | 29
10 | 27 14 | 14 6 | 11 6 |
| 3 | KNK | 15 | 12 | 6 | 6 |
| 4 | SIK, 58 | 8
14 | 12 16 | 8 15 | 6 10 |
| 5 | IAK, 46 | 13
14 | 15 14 | 11 10 | 8,0 8,0 |
| 6 | BPP | 18
17 | 16 16 | 12 14 | 9,0 7,0 |
| 7 | ChPN, 67 | 14
13 | 15 15 | 13 13 | 8,0 6,0 |
| 8 | DYG, 48 | 12
14 | 11,0 12,0 | 9,0 9,0 | 7,0 9,0 |
| 9 | FPO, 61 | 40
20 | 24 22 | 23 18 | 19 13 |
| 10 | UAP, 77 | 11
17 | 14 17 | 10 13 | 8,0 10 |
| 11 | FEI, 58 | 36
13 | 27 14 | 22 10 | 18 9,0 |
| 12 | OOO | 40
18 | 27 19 | 25 16 | 19 8,0 |
| 13 | TYB, 67 | 16
15 | 14 14 | 10 10 | 6,0 7,0 |
Analysis of Measurement Results
| | Var1 | Var2 |
Var3 | Var4 |
| MEAN | 18,52 | 17,16 | 13,24 | 9,84 |
| MEDIAN | 15 | 15 | 13 | 8 |
| SD | 9,06053714 | 5,24944442 | 5,13387443 | 4,2296572 |
| VALID_N | 25 | 25 | 25 | 25 |
| MIN case 1-61 | 8 | 11 | 6 | 6 |
| MAX case 1-61 | 40 | 27 | 25 | 19 |
| R | 0,87 | 0,61 | 0,52 |



While analyzing the results of IOP measuring using one and the same device by one and the same investigator, the tonometer position on the eyelid being changed, the mean values closeness was registered only when the position of tip on the eyeball was correct.
The sufficient correlation of the results was observed only in this case.
While moving the tip lower in the limb area or on the cornea the result is significantly lower and is not true.
|